Lesson 1Injection technique to minimise extravasation: vein selection, tourniquet use, aspiration check, flushing, and securing injection siteCovers IV injection methods to cut extravasation risk, including vein choice, tourniquet, aspiration, flushing, and site fixing, plus spotting, recording, and handling infiltration.
Patient assessment and vein selection criteriaTourniquet placement and vein visualizationAspiration check and slow, steady injectionSaline flush technique and line patency checksSecuring the injection site and patient positioningRecognizing and documenting infiltrationsLesson 2Contamination control and surveys: wipe tests, use of survey meters, frequency, acceptable limits, and action levelsAddresses contamination control in nuclear medicine, including regular surveys, wipe tests, limits, and response steps, stressing records, cleaning methods, and prevention.
Types of contamination and common sourcesUse and limitations of survey metersPerforming and interpreting wipe testsSurvey frequency and regulatory expectationsAction levels, decontamination, and re-surveyRecordkeeping and trend analysis for hotspotsLesson 3Dose calculations using time, activity, and patient weight: weight-based adjustments, decay correction, and paediatric/obese considerationsReviews dose calculation using time, activity, and weight, including weight dosing, decay fixes, and tweaks for kids and overweight patients within rules and limits.
Basic activity, time, and decay relationshipsUsing decay factors and half-life tablesWeight-based adult dosing calculationsPediatric dosing and minimum activity limitsAdjustments for obese or very small adultsChecking calculations against protocol rangesLesson 4Common radiopharmaceuticals and administration routes: Tc-99m sestamibi/tetrofosmin for myocardial perfusion (IV), Tc-99m MDP for bone scan (IV); typical adult activity rangesReviews common technetium-99m tracers, uses, routes, and adult activity ranges, stressing right choice, labelling, and records for heart and bone imaging.
Tc-99m sestamibi: indications and adult dose rangeTc-99m tetrofosmin: indications and adult dose rangeTc-99m MDP: indications and adult dose rangeIntravenous administration steps and checksContraindications and common adverse reactionsDocumentation of product, lot, and administered doseLesson 5Dose calibrator operation: daily background check, constancy test, geometry considerations, setting correct isotope and time-of-measurement correctionsDetails daily dose calibrator use, including background checks, constancy tests, geometry, isotope settings, and time corrections, stressing QC records and fixes.
Daily background and constancy proceduresQuarterly accuracy and linearity testing basicsGeometry dependence and volume considerationsSelecting correct isotope and calibration factorTime-of-measurement and decay correctionsQC documentation, trends, and troubleshootingLesson 6Activity verification steps: measuring drawn and administered activity, syringe technique, tare weight method, and documentationExplains verifying activity before and after giving using dose calibrator, including syringe methods, residuals, tare weights, and accurate patient records.
Pre-injection activity measurement in calibratorSyringe positioning and reproducible geometryMeasuring residual activity after injectionTare method for syringe and shield correctionsCalculating net administered activityRecording activity, time, and technologist IDLesson 7Post-injection precautions for patients: restroom use, hydration, discharge instructions, and breastfeeding guidanceExplains key post-injection advice, including toilet use, hydration, contact limits, and breastfeeding tips to cut radiation to family, public, and staff while keeping image quality.
Immediate restroom use and bladder emptyingHydration strategies to enhance tracer clearanceLimiting close contact with infants and pregnant womenWritten discharge instructions and key counseling pointsBreastfeeding interruption and milk storage guidanceWhen to escalate concerns to the nuclear medicine physicianLesson 8Safe handling and transport of radiopharmaceuticals: vial handling, needle/syringe best practices, use of lead shielding and dose traysDetails safe handling, packing, and moving of tracers, focusing on vials, syringes, shielding, labelling, and rules to cut exposure and avoid loss or errors.
Receiving, unpacking, and inspecting shipmentsVial handling, venting, and aseptic techniqueNeedle and syringe selection and secure assemblyUse of lead pigs, syringe shields, and dose traysLabeling, transport containers, and DOT categoriesSpill prevention during preparation and transportLesson 9Radiation protection principles (ALARA): time, distance, shielding applied to technologist workflow and room layoutApplies ALARA rules of time, distance, shielding to tech workflow, room setup, and patient scheduling, with practical tips to cut work dose and meet limits.
Time minimization during high-exposure tasksMaximizing distance using layout and toolsOptimizing shielding for staff and public areasRoom design, hot lab zoning, and traffic flowUse of dosimeters and review of exposure reportsPregnant worker considerations and declarations